Fluke Corporation MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 7-300: CAL ADJ /5500 INSTRUMENT: Fluke 7-600: CAL ADJ /5500 DATE: 28-Aug-96 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 58 CONFIGURATION: Fluke 5500A ============================================================================= # # Source: # Fluke 10 Series Service Manual (PN 900824 March 1992, Rev 1, 7/93). # # Compatibility: # 5500/CAL 4.0 or MET/CAL 4.0 or later # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5500A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 HEAD CALIBRATION ACCESS PROCEDURE 1.002 DISP Remove any input signals to the meter and turn off. 1.002 DISP 1.002 DISP Remove the four screws on the back and separate the 1.002 DISP case bottom and case top. 1.002 DISP 1.002 DISP Notice that when you do so the battery remains in the 1.002 DISP case bottom and power to the meter is disconnected. 1.003 DISP Observing correct polarity, connect a 9V battery to the 1.003 DISP battery contacts using easy hook jumpers or alligator 1.003 DISP clip leads. 1.003 DISP 1.003 DISP [32] NOTE 1.003 DISP To avoid stretching or bending the battery contacts, 1.003 DISP connect leads to the base of the contacts. 1.004 HEAD EQUIPMENT SETUP 1.005 DISP Ambient temperature: 18C - 28C. 1.005 DISP Relative humidity: less than 80%. 1.005 DISP Allow the UUT to stabilize. 1.006 HEAD CALIBRATION PROCEDURE 1.007 DISP Move switch to the center position. 1.007 DISP Press RANGE to select the 4V Range. 1.008 5500 4V S 2W 1.009 DISP Adjust R4 reading of 4.000+/-0.003V. 1.010 END